Description
TRICOR® Coriolis Mass Flow Meter TCM 230K for Gases and Liquids
MEASURING RANGES | |
Mass Flow Rate (max.): | 230000 kg/h, 8450 lb/min |
Volumetric Flow Rate (max.): | 230000 l/h, 1012 gpm, 34700 bbl/d |
PRESSURE | |
Standard Pressure Rating: | up to 100 bar, 1450 psi |
ACCURACY (LIQUIDS) | Stated flow accuracy combines the effects of repeatability, linearity and hysteresis |
Basic Accuracy [Mass flow]: | ±0.1 % of flow rate |
Mass Flow Repeatability: | ±0.05 % of flow rate |
Basic Accuracy [Volume flow]: | ±0.15 % of flow rate |
Volume Flow Repeatability: | ±0.05 % of flow rate |
Zero Stability: | ±0.01 % of full scale |
ACCURACY (GASES) | |
Basic Accuracy [Mass flow]: | ±0.5 % of flow rate |
Mass Flow Repeatability: | ±0.25 % of flow rate |
DENSITY | |
Density Range: | up to 2500 kg/m³, 2.5 g/cm³ |
Density Accuracy: | Liquids ±1.0 kg/m³, ±0.001 g/cm³ |
Density Repeatability: | ±0.5 kg/m³, ±0.0005 g/cm³ |
TEMPERATURE | |
Process Temperature Range: | -40 °C … +100 °C, -40 °F … +212 °F |
Temperature Accuracy: | ±1 °C ±0.5 % of reading ±1.8 °F ±0.5 % of reading |
Temperature Repeatability: | ±0.2 °C, ±0.36 °F |
MATERIALS OF CONSTRUCTION (wetted parts) | |
Tubes: | 1.4404/AISI 316L |
Case: | 1.4404/316L |
Flow splitter: | CF3M/1.4409/AISI 316L |
Brazing Alloy: | BNi-2 |
EX CERTIFICATIONS | |
ATEX (Ex) | Zone 1: Group IIC or IIB, T2…T4 |
ATEX (Exn) | Zone 2: II 3G Ex nA IIC T2…T4 Gc |
IECEx (Ex) | Zone 1: Group IIC or IIB, T2…T4 |
cCSAus (Ex1) | Class 1, Division 1: Group A, B, C, D or C, D, T2…T4 |
ATEX + IECEx + cCSAus triple approval (Ex3) | Zone 1: Group IIC or IIB, T2…T4 and Class 1, Division 1: Group A, B, C, D or C, D, T2…T4 |
Only with remote electronics | |
For cCSAus and triple Ex-approval the Ex-approvals for remote electronics in 1.4404/316L is pending | |
EAC (TR-CU) | Group IIC or IIB, T2…T4 |
For cCSAus and triple Ex-approval the Ex-approvals for remote electronics in 1.4404/316L is pending |